6.4 Disinfection
Disinfection must be performed no later than 2 hours after the cleaning phase.
Automated disinfection is preferred if conditions permit.
Automated disinfection-Washer-disinfector
·The washer-disinfector is proved to be valid by CE certification in accordance with
EN ISO 15883.
·Use high temperature disinfection function. The temperature does not exceed 134 ° C,
and the disinfection under the temperature cannot exceed 20 minutes.
·The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883.
Cleaning and disinfecting steps by using Washer-disinfector:
1.Carefully place the product into the disinfection basket. Fixation of product is
needed only when the product is removable in the device. The products are not
allowed to contact each other.
2.Use a suitable rinsing adaptor, and connect the internal water lines to the rinsing
connection of the washer-disinfector.
3.Start the program.
4.After the program is finished, remove the product from the washer-disinfector,
inspect (refer to section "Inspection and Maintenance") and packaging (refer to
chapter "Packaging"). Dry the product repeatedly if necessary (refer to section
"Drying").