19
Problems
Possible causes
Solutions
If the root canal is filled with blood,
liquid?
Once the endo needle contact the surface
of the root canal which is filled with
blood, liquid, it will display “OVER”
immediately. In this case, push the needle
to the apical root canal, then the display
will be normal, you can measure the
length of the root canal correctly.
If there is liquid, scrap on the tooth
surface?
Clean the tooth surface.
If the endo needle contact the gums?
It will display “OVER” if the endo needle
contact the gums.
Display not steady while
measuring: the measurement
result is rather longer or
shorter; numerical display
irregular.
If there is still pulp in the root canal?
If there is much pulp left in the root canal,
the root canal length can’t be measured
correctly.
If the needle touched the metal repaired
material?
Once the needle touched the metal
repaired material, current measurement
from the gums to the periodontal tissue
loss, the screen will display “OVER”.
If the adjacent surface has caries?
Current measurement flow from caries of
the adjacent surface to gums, then the root
canal length can’t be measured correctly.