BLUETOOTH PRINTING ERRORS
BT Module Not Found!
Interpretation:
No response from the internal Bluetooth Module.
Response:
Try again. If still no response contact Greyline Instruments.
Failed to connect to printer!
Interpretation:
The unit has failed to make connection with the
default
Bluetooth
printer.
Response:
Try again. If connection still fails then start a search
for
another
printer.
4-20mA ERRORS
mA out > Max
Interpretation:
The actual flow is higher than the maximum set on the mA range.
Response:
Re-scale the 4-20mA output to be able to cope with the higher flow –
see
Paragraph 3.7.4
.
Calibration 20mA Error!
NOTE: The 4-20mA output is calibrated before the instrument leaves the factory
and should not require further adjustment
.
Interpretation:
You have adjusted the DAC outside its accepted range when
calibrating the 20mA signal output.
Response:
Re-calibrate the 4-20mA output – see
Paragraph 3.7.4
.
Calibration 4mA Error!
NOTE: The 4-20mA output is calibrated before the instrument leaves the factory
and should not require further adjustment.
Interpretation:
You have adjusted the DAC outside its accepted range when
calibrating the 4mA signal output.
Response:
Re-calibrate the 4-20mA output – see
Paragraph 3.7.4
.
DATA LOGGING ERRORS
Log not empty!
Interpretation:
Not yet implemented.
Log memory full
Interpretation:
This occurs when all the data logger memory locations are filled.
The effect on the logging process will depend on the setting of the Memory rollover
field in the REAL TIME LOGGER screen (which may be set to Stop or Overwrite).
Response:
Clear the logger memory, as described in
Paragraph 3.6.3
.