Page
21 www.greyline.com
site parameters will be listed on the screen.
5. Scroll down through the menu list and
enter/change the data that might have changed
since the last time the site was accessed.
6. When you are satisfied that the parameters are
correct select Save current site & read flow.
Pipe wall material : Mild Steel
Lining material
: ---------
Sensor set
: A-ST
Sensor mode : Reflex
Fluid type
: Water
Save current site & read flow
Delete this Site
Download & save current site
Exit
FLUID TEMPERATURE
DD-MM-YY HH:MM:SS
Enter Fluid Temperature
°C: 5.00
°F: 41.00
Continue..
7. If you need to alter the fluid temperature from that
shown select either °C or °F with the cursor and
press the ENTER key.
8. Enter the new temperature value and press the
ENTER key.
9. The new temperature should now be indicated in
both °C and °F.
10. Select Continue.. and press ENTER.
SENSOR SEPARATION
DD-MM-YY HH:MM:SS
Site
: MySite
Pipe
: 58.00 mm
Wall
: 4.00
Sensors
:
A-ST
Reflex
Temperature
: 10.00°C 50.00°F
Set sensor separation to
32.2 mm
Press
to continue,
to select sens.
11. The SENSOR SEPARATION screen now displays
a summary of the entered parameters and informs
you of the type of sensor to be used, the mode of
operation and the distance to set up between the
sensors.
In this example it recommends type A-ST (A
standard) sensors operating in the ‘Reflex’ mode
spaced 32.2mm apart.
Take a note of these details.
Note:
If you press ENTER before the transducers are fitted and connected to the instrument, the instrument
will detect a low signal strength and may indicate an ERROR condition.
Attaching and connecting the transducers
12. Fit the designated sensors to the pipe using the appropriate guide rails as described in
Paragraph 2.2
. Take
great care to set the separation distance as accurately as possible.
13. Connect the red and blue coaxial cables between the sensors and the instrument, ensuring that the red
connector on the instrument is connected to the ‘upstream’ sensor.
Taking a flow reading
14. Once the transducers have been fitted and
connected press the ENTER key.
15. This will take you from the SENSOR SEPARATION
screen to the FLOW READING screen via a signal-
checking screen (shown here).
Please wait..
Checking signals
****************************************
*
*
****************************************