BATTERY ERRORS
Battery Low
Interpretation:
The battery has discharged to below 30% remaining. This leaves
the instrument with approximately 4 hours remaining, depending on power usage,
before it needs recharging.
Response:
Recharge the internal battery at the earliest opportunity. Do not leave
the instrument for a prolonged period with a fully discharged battery.
Battery Exhausted
Interpretation:
The battery is approaching a fully discharged state and the
instrument is about to store the internal data and shut-down.
Response:
Recharge the battery.
SET-UP ERRORS
Pipe OD out of range
Interpretation:
You have entered an out-of-range value for the pipe outside
diameter dimension – i.e. larger or smaller than the unit or sensor can be used on.
Response:
Enter a valid number.
Wall thickness out of range
Interpretation:
You have entered an out-of-range value for the pipe wall thickness
dimension – accepted range is 0.04” - 2.95” (1mm - 75mm).
Response:
Enter a valid number.
Lining thickness out of range
Interpretation:
You have entered an out-of-range value for the lining thickness
dimension – acceptable range is 0” – 0.39” (0mm - 10mm).
Response:
Enter a valid number.
Temperature range
Interpretation:
You have entered an out-of-range value for the fluid Temperature.
Accepted temperature range -4°F to 572°F (-20°C to +300°C).
Response:
Enter a valid number.
Invalid Date or Time
Interpretation:
The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.
Response:
Enter a valid Date and Time.
Sensors: INVALID
Interpretation:
The selected temperature is higher than the maximum allowed for
the sensor type.
Response:
Select alternative sensors or change the temperature.
Mode: Err Typ
Interpretation:
The selected sensors are invalid and the mode cannot be verified.
Response:
Select a valid sensor type and choose a mode that gives a non-zero
separation distance.