-67-
Guidance and manufacturer’s declaration - electromagnetic emissions
The RigiScan is intended for use in the electromagnetic environment specified below. The customer or the user of
the RigiScan should assure that it is used in such an environment.
Emissions
Test
RF emissions
CISPR 11
The RigiScan uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The RigiScan is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Group 1
Group B
TBD
TBD
RF emissions
CISPR 11
Harmonic
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment — guidance
Guidance and manufacturer’s declaration - electromagnetic immunity
The RigiScan is intended for use in the electromagnetic environment specified below. The customer or the user of
the RigiScan should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61.000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
± 6 kV contact
± 8 kV air
3 A/m
3 A/m
±2 kV for power
Supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to
line(S)
± 2 kV line(s) to earth
<5 % (IT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in U-1)
for 5 cycles
70 % (IT
(30 % dip in UT)
for 25 cycles
<5 % UT 180°
(>95 % dip in UT)
for 0.5s
<5 % (IT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in U-1)
for 5 cycles
70 % (IT
(30 % dip in UT)
for 25 cycles
<5 % UT 180°
(>95 % dip in UT)
for 0.5s
± 1 kV line(s) to
line(S)
± 2 kV line(s) to
earth
±2 kV for power
Supply lines
± 1 kV for
input/output lines
Compliance
Electromagnetic environment
– guidance –
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
RigiScan requires continued operation during power
mains interruptions, it is recommended that the
RigiScan be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.
Содержание RigiScan Plus
Страница 1: ...User Guide CAUTION Federal USA law restricts this device to sale by or on the order of a physician...
Страница 60: ...60 Appendix A Product Specifications Physical Dimensions RIGISCAN User Guide...
Страница 70: ...70 RIGISCAN User Guide Appendix E RigiScan Nomogram...
Страница 72: ...72 RIGISCAN User Guide NOTES...
Страница 73: ...73...