6. Tighten the plastic retention nut on the tension guide until it contacts the Tension Guide housing.
(see Figure 4-1)
4.2 Removing the Tension Guides
1. Verify that the RigiScan Plus Monitor is turned off.
2. Unscrew the plastic retention nut on the Tension Guide at
the monitor end of the Tension Guide.
3.
Expand the loop to its maximum circumference by inserting
your fingers in the loop and slowly applying outward pres
-
sure until the loop will no longer expand. (see Figure 4-2)
Fig. 4-1 Installing the Tension Guides
Fig. 4-2 Expanding the Loops
CAUTION
: DO NOT over tighten the retention nuts as this will make future removal very
difficult.
CAUTION
: Turn the RigiScan Monitor “off” before replacing or removing the Tension Guides/
cables. Operating the Monitor without the Tension Guides/cables in place may damage the de-
vice or result in inability to insert the Tension Guides. Care must be taken to remove the tension
guides slowly and gently to avoid damage to the cables or internal mechanisms. Removing the
Tension Guides too quickly may damage the monitor or result in an ‘out-of-phase’ condition and
inability to re-insert the tension guides.