-6-
R
IGI
S
CAN
®
(
User Guide
)
electrocautery, electrocoagulation, diametry or telemetry equipment, high power transmit-
ters, or household appliances such as electric blankets, electric razors, power tools, micro-
wave ovens, hair dryers or cellular phones. Items such as these may interfere with operation
of the monitor and cause erroneous readings.
• Patients MUST NOT bath or shower while wearing the RigiScan Plus Monitor. Getting the
monitor wet may cause electrical shock to the patient or damage to the monitor.
• Patients MUST NOT engage in any sexual activity, such as intercourse, oral sex or mastur-
bation, during RigiScan Plus testing. Doing so may result in patient injury or erroneous data.
• Patients with impaired penile sensation are at greater risk of exposure to maximum compres-
sive forces should the system fail to operate properly.
• DO NOT turn the RigiScan Plus Monitor “on” until the loops have been placed on the penis.
Turning the monitor on before placing the loops will cause continuous rigidity sampling and
may result in patient discomfort, pain, irritation or bruising.
• The RigiScan Monitor must be turned “off” prior to urination. The monitor may be turned
off for a continuous period of up to 15 minutes without interrupting the monitoring session.
Urinating with the monitor on may cause pain, urethral irritation or hematuria.
• The RigiScan Monitor must be turned “off” prior to readjusting the position of the loops on
the penis. Adjusting the loops with the monitor on may cause continuous rigidity sampling
and result in patient discomfort, pain, irritation or bruising.
• Reuse of disposable loop covers may result in patient injury or transmission of communica-
ble disease.
• When the RigiScan Plus System is to be used in a Real-time Monitoring Mode (RTM), the host
PC must use an isolation transformer manufactured to EN60601 standards. Failure to do so
may result in a safety hazard.
Cautions
•
Ingestion of coffee, tea, cola, alcoholic beverages, sedatives, tranquilizers, muscle relaxants
or sleeping pills in the period immediately preceding nocturnal testing with the RigiScan Plus
may affect test results by interfering with normal sleep patterns.
Warnings
(continued)
Содержание RigiScan Plus
Страница 1: ...User Guide CAUTION Federal USA law restricts this device to sale by or on the order of a physician...
Страница 60: ...60 Appendix A Product Specifications Physical Dimensions RIGISCAN User Guide...
Страница 70: ...70 RIGISCAN User Guide Appendix E RigiScan Nomogram...
Страница 72: ...72 RIGISCAN User Guide NOTES...
Страница 73: ...73...