GMC-I Messtechnik GmbH
23
dLY – Output Off Delay for OCP
Functions
•
Delay time prior to deactivation of the power output after
current regulation has been triggered (
Iout = Iset)
•
Only enabled with activated
OCP
function (
OCP ON
)
•
If output current
Iout
drops below
Iset
before
DELAY
time
elapses, the shutdown sequence is aborted.
•
If current regulation is triggered again, the routine is started
over (at 00.00).
•
The default setting after
RESET
(
*RST
) is 00.00.
Settings
4.6 Display of Momentary Output Values Uout, Iout
and Pout
Uout – Momentary Measured Voltage Value
•
Appears at the
left-hand display
[5].
•
The
green Uout/V
LED [6] lights up and indicates display of the
momentary measured voltage value.
•
The measured voltage value is automatically displayed again,
approximately 10 seconds after the last setting has been
made.
•
The momentary measured voltage value is displayed
immediately if the user exits the device functions setting mode
by pressing the
<CE/LOCAL>
key [16].
Iout – Momentary Measured Current Value
•
Appears at the
right-hand display
[8].
•
The green
Iout/A
LED [9] lights up and indicates display of the
momentary measured current value.
•
The measured current value is automatically displayed again,
approximately 10 seconds after the last setting has been
made.
•
The momentary measured current value is displayed
immediately if the user exits the device functions setting mode
by pressing the
<CE/LOCAL>
key [16]
Pout – Display Momentary Output Power
Calculated internally based upon momentary measured values
Uout and Iout.
Settings
See description in chapter 4.8.1.
UI
_
– Display Measured Values in U/I Min-Max Memory
Functions
•
Reads out stored values for
Umin
,
Umax
,
Imin
and
Imax
at the
display, or via the computer interface.
•
Stored Min-Max values can be read out regardless of the
status selected for the
MINMAX
function.
Settings
See description in chapter 4.8.1.
4.7 Operating Menu via the FUNCTION Key
The
FUNCTION
menu consists of the following functions for
configuring the KONSTANTER’s parameters:
Function group
Function
Parameter (numeric / text)
■
Setup
SEt
Ulim
NP: xx.xx
Ilim
NP: xx.xx
OVP
NP: xxx.x
OCP
TP:
oFF
/ on
dLY
NP: xx.xx
Pon
TP: rSt / SbY / rcL
UI
-
TP:
oFF
/ on / rSt
rnd
TP:
0
/ -1 / -2
■
Analog Interface
AnIF
trG
TP: oFF / out / rcL / SEq / LLO / UI
-
SiG1
TP: oFF / on / out / Mode / SEq /
SSEt / U-Lo / U-Hi / I-Lo / I-Hi
SiG2
TP: oFF / on / out / Mode / SEq /
SSEt / U-Lo / U-Hi / I-Lo / I-Hi
■
Sequence Function
SEq*
tSEt
NP: xx.xx
SSEt
NP:
oFF
/ on
tdEF
NP: xx.xx
Strt
NP: xxx (11... 252)
StoP
NP: xxx (12... 253)
rEP
NP: xxx (cont. or 1 ... 255)
SEq**
TP: Go, Strt, StoP, hold, StEP, cont.
* Sequence function: appears at left-hand display, right-hand display is blank.
** Parameter selection for sequence control: appears at left-hand display, selected parameter
appears at right-hand display.
■
Interface Configuration
buS
Addr
NP: 0 / 1 / ... /
13
/... / 30 / (UNL)
bAUd
NP: 50 / 75 / 150 / 200 / ... / 4800 /
9600
/ 19200
dbit
NP: 7 /
8
Pbit
TP:
nonE
/ ZEro / EVEn / odd / onE
Sbit
NP:
1
/ 2
■
Adjustment (calibration, chapter 7)
CAL
dAtE
TP: mm.yy
last/new adjustment date
CAL
TP: Strt
MEAS
UoFF
UoFF
NP: x.xxx
entry of ext. measured value U
MEAS
UFS
UFS
NP: xx.xx
entry of ext. measured value U
MEAS
IoFF
IoFF
NP: x.xxx
entry of ext. measured value I
MEAS
IFS
IFS
NP: xx.xx
entry of ext. measured value I
CAL
End
NP: Numeric Parameters
•
Can be selected as a specified value or as a value within an
interval with corresponding resolution.
•
Adjust
resolution
(decimal place) by pressing the
<SELECT>
key.
•
Select with the
<
>
key or the
<
>
key.
•
Displayed value = selected value.
TP: Text Parameters
•
Select parameter with the
<
>
key or the
<
>
key.
•
Acknowledge the selected value by pressing the
<ENTER>
key.