OCP – Overcurrent Protection
Functions
See functions description on page 22.
Setting Parameter
☞
OFF: OCP
function
is inactive
Continuous current limiting (current regulation)
☞
ON: OCP
function
activated
The output is deactivated as soon as current limiting has been
active for the specified
DELAY
time.
☞
The selected text parameter does not become active until the
<
ENTER
> key is pressed!
Settings
The procedure is described in principle in chapter 4.7.
!
The
OCP
display and a related
text parameter appear in the
setup function window.
Figure 4.7.1 b
Path to Selection of the OCP Text Parameter
Note
Load current for the output capacitor is also acquired by the
current regulator, and is limited to a value of Iset
Ioad. If the
OCP
function has been activated and
ISET
and
DELAY
have been set to
low values, the output may even be deactivated if output voltage
is increased. For this reason,
DELAY
must at first be set somewhat
higher than the resulting output voltage response time.
dLY – Output Off Delay for OCP
Functions
See functions description on page 23.
Settings
The procedure is described in principle in chapter 4.7.
!
The
DELAY
display appears in
the window for the setup
function, along with a related
numeric parameter [t].
Pon – Specifying a Power-On Status
Functions
•
Use automatic KONSTANTER configuration after power on.
Setting Parameters:
☞
rSt: (RESET
) Factory default settings are utilized
☞
SbY: (STANDBY
) Same settings as prior to shutdown, power
output remains inactive (
OUTPUT OFF
)
Disabled panel controls are re-enabled after mains ON.
From firmware version 4.003 onwards:
Disabled panel controls remain blocked even after mains ON.
☞
rcl (RECALL
) Same settings as prior to shutdown.
Disabled panel controls remain blocked even after mains ON.
Settings
The procedure is described in principle in chapter 4.7.
!
The
Pon
display and a related text parameter appear in the
setup function window.
Figure 4.7.1 c
Path to Selection of the Pon Text Parameter
Note
•
The status of the
Pon
function is not saved as a device setting
to
SETUP
memory with the
<SAVE>
key.
•
The following settings are recommended, depending upon
how the KONSTANTERs is used:
!
Pon rSt
:
Use in computer controlled systems
!
Pon SbY
:
Common laboratory use
!
Pon rcL
:
Applications which must continue in an
unchanged fashion after power failures.
*
Select the text parameter
from an options loop
From the SEt function group
ENTER
CE/LOCAL
FUNCTION
*
*
Select the text parameter
from an options loop
From the SEt function group
ENTER
CE/LOCAL
FUNCTION
*
Содержание SSP 120 Series