38
GMC-I Messtechnik GmbH
Figure 4.7.3 n
Suspending and Resuming a Sequence
Executing the Next Address and Beginning Step-by-Step Control
(status: HOLD
HOLD)
Functions
!
The next memory location can be selected at any time from a
suspended
SEQUENCE
. The setpoints saved to the next
memory location are executed. The selected dwell time of the
exited and the new memory address is not taken into
consideration.
!
The device is switched from automatic sequence to step-by-
step control the first time this step is executed (see page 36).
•
The output retains its switching status (output On or Off).
•
When jumping from the start address to the stop address, the
remaining repetitions counter (rrep) remains unchanged.
Procedure:
☞
With
suspended
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
> key, the <
> or the <FUNCTION> key
until SEq
step
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
executes the next address.
!
The selected dwell time of the new and the exited memory
address is not taken into consideration.
!
The LOCKED/SEQ LED blinks slowly in the HOLD status
.
Executing the Start Address and Beginning Step-by-Step Control
(status: HOLD
HOLD or RUN
HOLD)
Functions
!
The start address can be selected at any time from a
suspended or a running
SEQUENCE
. The setpoints saved to the
start address are executed. The selected dwell time of the
exited and the new memory address is not taken into
consideration.
!
The device is switched from automatic sequence to step-by-
step control the first time this step is executed (see page 36).
•
The output is switched to the output on status.
•
The remaining repetitions counter is reset to the initial value
(rrep = rep).
Procedure:
☞
?
With
interrupted
or
running
SEQUENCE,
– press the <FUNCTION> key and then browse with the
<
> key, the <
> or the <FUNCTION> key
until SEq
start
appears
and execute the command with the
<ENTER>
key.
!
The
SEQUENCE
executes the start address.
!
The selected dwell time of the new and the exited memory
address is not taken into consideration.
!
The LOCKED/SEQ LED blinks slowly in the HOLD status
.
Adding New Settings to an Existing SEQUENCE
Deleting an Individual SEQUENCE Memory Location
4.7.4
buS – The “Interface” Function Group
The KONSTANTER can be set up for data transmission to a
computer in the interface configuration function group. In this
case, all of the settings previously selected with DIP switches are
entered via menu functions.
The serial port (RS 232) or the (optional) parallel port (IEEE 488) is
used for communication.
Addr – Setting the Device Address
•
An address must be assigned to the SSP KONSTANTER in
order to assure that it functions correctly together with other
devices on the IEC bus.
•
If the KONSTANTER receives its own address from the computer, it
interprets subsequent data and executes the respective commands.
•
If it receives the address of another device, all subsequent
data are ignored until it receives its own address.
Settings
☞
Procedure
for selecting
Addr
, and setting the desired parameter
☞
Press the <
FUNCTION
> key in order to access the
FUNCTION
menu.
!
The last edited function or function group is accessed
automatically.
☞
Select
buS
configuration if necessary by pressing the
<
FUNCTION
> key, or the <
FUNCTION
> key and simultaneously
the <
> key or the <
> key.
!
buS
appears at the left-hand display.
☞
Acknowledge by pressing the <ENTER> key.
ENTER
ENTER
ENTER
SEQUENCE
SEQUENCE
Sequence is resumed
Resume sequence
Restart sequence
End sequence
Select:
The setting
is executed
in hold status
LOCKED/SEQ
LOCKED/SEQ
in run status
LOCKED/SEQ
Step-by-step, start address
Step-by-step, next address
FUNCTION
FUNCTION
CE/LOCAL
End at
current address