SiG1 – Signal Output 1
Functions
•
Digital open collector output with reference to AGND, max.
switching voltage min. 30 V DC, max. switching current min.
20 mA (e.g. ext. load shedding relay).
Indicates the status of the power output.
•
If an event occurs which is associated with the selected
parameter, a corresponding signal is generated at the output.
•
Used to link the OUTPUT function via the trigger input in the
case of parallel or series (master-slave) connection of several
SSP KONSTANTERs.
•
Control of an external output relay is possible.
•
As a status signal for monitoring devices
•
As a freely assignable switching output
1
The
SSEt
parameter must be set to
ON
for a certain memory location in the
SEQUENCE
function
group.
☞
When the
SEQUENCE
is run, an signal is additionally generated
for the read-out of Uset and Iset when this particular memory
location is executed.
☞
The signal is available at the outputs of the analog interface
(SiG1 / SiG2 / SiG1 + SiG2) for the respectively selected dwell
time
tSEt
or
tdEF
.
Setting
☞
Procedure
for selecting
Sig1:
see also
chapter 4.7, “Settings”.
!
The
Sig1
display and a related text
parameter (status) appear in the
AnIF
function window.
Figure 4.7.2 b
Path to Selection of the SiG1 Text Parameter
SiG2 – Signal Output 2
Function and Parameters
☞
Identical to
SiG1
Settings
☞
Identical to
SiG1
Setting Parameters
SIG1 OUT Status
☞
OFF
Always passive
high
☞
ON
Always active low
☞
OUT
Power output:
OUTPUT ON
OUTPUT OFF
Passive high
Active low
☞
ModE
Control mode signal:
Constant current regulation
Overload (Pmax)
Constant voltage, output Off
Active low
Active low
Passive high
☞
SEq
Status with running sequence
Active low
☞
SSEt
If
SSEt
is set to ON for special
memory location in the
SEQUENCE
1
,
otherwise
Active low
Passive high
☞
U_LO
Umeas < Uset
(reference value memory
Umeas
Uset
254
)
Active low
Passive high
☞
U_HI
Umeas > Uset
(reference value memory
Umeas
Uset
255
)
Active low
Passive high
☞
I_LO
Imeas < Iset
(reference value memory
Imeas
Iset
254
)
Active low
passive high
☞
I_HI
Imeas > Iset
(reference value memory
Imeas
Iset
255
)
Active low
Passive high
Select the text parameter
from an options loop
From the AniF function group
*
ENTER
CE/LOCAL
FUNCTION