Naris
Naris
User’s
Manual V14 14/02/20 Page 41 of 52
NR6 RHINOMANOMETER
CALIBRATION
A calibration check should be made, and if required the instrument adjusted, in the following
circumstances:
a)
if the pressure or flow transducers are changed
b)
if the flowhead is contaminated with dust or other particles or has been washed or
disinfected
c)
if some time has elapsed since the last calibration (ideally a check should be made
before each testing session)
d)
if there is any uncertainty about the results achieved
Two transducers are used in each NR6 - one for measuring the flow and one for measuring the
pressure. They can be set using static individual calibration units, such as a liquid column for
pressure and a rotameter for flow.
CALIBRATION EQUIPMENT
to NR6 Pressure Port
(Black)
A.
Pressure Gauge
- a liquid column
covering the range 0-500 Pa (51mm H
2
0).
It is an added convenience if this is
connected in the manner shown because
adjustment to the desired pressure level
is more easily achieved
Hand Pump Reservoir Liquid
Column
B.
Flow Gauge
- such as a rotameter
covering the range 0-500cm3/sec.
(0-30 l/min) and an adjustable
Rotameter
source of flow with a variable
Vacuum
voltage supply. Such a motor is
Motor
Flowhead
available from manufacturers of
vacuum cleaners and the speed at
which it runs, and therefore the flow
it produces, can be controlled by
powering it from a variable trans- Variable former
Transformer
to NR6
The procedure to be employed is as follows: