Naris
Naris
User’s
Manual V14 14/02/20 Page 10 of 52
Standardisation Document
Customers are referred to the under noted International Standardisation Committee
publication which contains recommendations on the use of Acoustic Rhinometers.
Acoustic Rhinometry: Recommendations for technical specifications and standard
operating procedures.
Rhinology supplement number 10, December 2000
Ole Hilberg and Ole Pedersen
A1 ACOUSTIC RHINOMETER TECHNICAL SPECIFICATIONS
Only factory-trained personnel or engineers who are familiar with the standard EN 60601
can undertake servicing of the A1 Acoustic Rhinometer.
Circuit diagrams will be made available to competent persons on request.
Medical CE Mark
The CE mark indicates that the device meets the requirement of
Annex V & VII of the Medical Device Directive 94/42/EEC
Standards
Safety
BE EN 60601-1 :2006
EMC
BS EN 60601-1:2007
Performance
(using standard nose)
Repeatability
Better than 2%
Volume
Accuracy (0cm
to 5cm)
Better than 2%
Minimum Area
Accuracy
(within 0cm to
5cm)
Better than 5%
Area Range
(max 20,
default 10)
0.1 cm
2
to 10 cm
2
Distance Range
(using standard
sound tube)
0 cm to 12 cm
Calibration
Self - calibrating
Operation Environment
Temperature
+15
o
C to +35
C
Relative
Humidity
20% to 80% RH non-condensing
Pressure
50 kPa to 106 kPa
Duty Cycle
Continuous
Warm Up Time
5 minutes
Supply
Universal voltage
–
external supply
Power
10 Watts
Transportation and Storage
Temperature
-40 °C to +60 °C
Relative
Humidity
20% to 80% RH non-condensing
Pressure
50 kPa to 106 kPa
Mechanical Performance
Size
27 cm x 8 cm x 30 cm
Weight
2 Kg