APPENDIX 2: TECHNCIAL INFORMATION (relative to EN 60601-1:2006)
No modification to this equipment is allowed
.
6.2
When the NR6 Rhinomanometer is powered by a PC, power can be removed by
unplugging the PC mains plug, by switching power off at the mains plug, or by removing the
USB cable from either socket.
It complies with the specification for Class I ME equipment, and can only be used along with
a PC and printer, normally supplied by the customer. NB the warning made earlier about the
PC and printer being compliant with EN 60950 and placing these items out of the reach of a
subject.
The Applied Parts comprise, mask, anti-viral filter or posterior or anterior tubing, tip
connector, microfoam tape, foam inserts, flowhead and silicone rubber tubes which are
classified as type B applied part.
6.3
The enclosure does not provide protection against the ingress of liquids and is therefore
classified as IPX0.
6.4
There are no parts which are supplied sterile or are required to be sterile.
6.5
The NR6 Rhinomanometer is
NOT
suitable for use in an oxygen rich environment.
6.6
The NR6 Rhinomanometer rated for continuous use.
7.2.2
The NR6 serial number can be found on the label on the back panel, and the software
version number can be viewed on the start-up screen and by clicking on Help, found on the
top bar of the NR6 software
7.2.3
Information is given in the Warning and Caution section of this manual, which the user
must read prior to operating this instrument. Appropriate labels are fixed to the back panel
7.2.4
Single use parts are identified by packaging labels detailing single use only
7.2.5
NR6 cannot be powered directly from the mains supply
7.2.6
The NR6 USB connection provides DC voltages of 5 volts dc.
7.2.11
The NR6 Rhinomanometer is rated for continuous use.