Manual V14 14/02/20 Page 21 of 52
ANTERIOR TEST PROCESS
Start the NR6 software, and set up a new record for a new patient, or click on
Open
if the
subject already has a record file. Click on the NR6 Acquire button to access the recording
screen (flow/pressure grid) then:
Insert the foam to the reference nostril
Position the tube on the soft part of the cheek
Test for leakage
Place the mask on the face to secure the tube
Ask the subject to quietly breath through their nose with their mouth closed. If the traces reach
the default sample point, click on the red button to record data. If not ask them to increase the
depth of their breathing. After four cycles have been recorded (default), move the foam (or
tape) to the other nostril and repeat the process.
ANTERIOR COMPONENTS
Hole Punch
Anterior Tape
Anterior tip connector Single Use Mask
Anterior mask with Tape/Tip connections
Anterior mask with Foam connections