Naris
Naris
User’s
Manual V14 14/02/20 Page 13 of 52
INSTALLATION
SOFTWARE
Installation
of the NR6 Software is described fully in the accompanying manual titled “
NARIS
Software Manual”
INSTALLING THE NARIS HARDWARE
NR6 HARDWARE CONNECTIONS
Please see the pictures below for the NR6 hardware connections:
NR6 front view - flowhead and
Back panel view showing a link
pressure connections cable socket to A1 and USB socket
The NR6 box has four nozzles on it, which are colour coded: Black is the pressure input, Blue
is the pressure reference, Red is the positive flow input, Green is the negative flow input.
Connect each tube onto the input nozzle with the corresponding coloured marker, i.e. black
-> black, red -> red etc.
Connect the free ends of the tubes to their corresponding coloured markers
–
green and red
tubes onto the green and red nozzles on the flowhead, and the tube marked blue onto the
mask port marked blue. If you are going to be making posterior tests, the tube marked black
will also go onto the mask nozzle marked black, but if you are going to make anterior
measurements, it will go onto an anterior tube connector. In this case the mask nozzle
marked black should be closed, using the plug provided
A1 HARDWARE CONNECTIONS
SOUND
TUBE
SOCKET
POWER SOCKET
FROM PCM5OUT04