InstaCal
Navigate to the MCC (Measurement Computing Company) folder, using
My Computer
and
double-click on the program
Inscal32.exe
- C:/Program Files(x86)/Measurement
Computing/DAQ.
N.B.
In a 32-bit system this will be C:/Program Files/Measurement Computing/DAQ as a 32-
bit system won’t have
an x86 folder.
The opening dialog will show which boards are installed.
•
The board you installed should be numbered 0 (or the corresponding number in the
A1 software, if for some reason you have multiple devices installed)
•
Select
Test
from the top bar and
Analog
from the drop-down menu then
CH0
, signal
source
External
and A/D range
+ 10V (for 201 USB interface)
The line you see should lie on zero Volts and should move when talking into the
open end of the sound tube
–
if this is ok then the microphone and connections
between A1, the A/D card and the PC are ok.