GlideScope
®
Cobalt AVL System
Quick Reference Guide
11
2. Disconnect the video baton from the monitor.
3. Place the protective cleaning cap over the connector. Make sure that the
mark on the cap is aligned with the arrow on the cable.
4. Wash the video baton manually using a hospital grade equipment
detergent or an enzymatic debridement agent/detergent to remove all
foreign material (eg, soil and organic material) from the surface of the
device.
5. The video baton can now be disinfected.
Correct cleaning position –
protective cap covering the
electronic connector
Cleaning the GlideScope
Cobalt AVL Video Baton
The GlideScope Cobalt AVL Video Baton, which is protected by the GVL
®
Stat, is not intended to have direct patient contact.
Caution
Do not place the GlideScope Cobalt AVL Video Baton in the cradle
if it is contaminated.
The video baton is a nonsterile, resuable device, which is protected from
direct contact with the patient by the sterile single-use Stat. The video baton
will normally only require cleaning and wiping down between uses when
used as intended. High Level Disinfection is required for the video baton
when it is visibly soiled.
To Clean the Video Baton
1. Detach the GVL Stat from the video baton.
Note: A used GVL Stat is a biohazard and must be disposed of in a
manner consistent with local protocols.
Содержание Cobalt AVL
Страница 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Страница 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Страница 26: ......
Страница 27: ...GlideScope Cobalt AVL System User s Manual...
Страница 28: ......
Страница 29: ...GlideScope Cobalt AVL System User s Manual...
Страница 30: ......
Страница 77: ......