GlideScope
®
Cobalt AVL System
User’s Manual
41
Cleaning the Cobalt AVL
Disinfecting the GlideScope
Cobalt AVL Video
Baton
To disinfect use any of the chemicals listed in Table 3 and do the following:
1. Ensure the equipment is clean (see Cleaning Section for video baton).
2. Ensure the protective caps on the connectors are secure.
3. Prepare the disinfectant solution at the concentration and temperature
recommended by the disinfectant manufacturer. Disinfect the equip-
ment following the disinfectant manufacturer’s instructions or as stated
in Table 3 on page 37.
4. After the disinfection process, rinse (as applicable) and dry the video
baton and then store the equipment in a clean environment.
Cleaning the GlideScope
Cobalt AVL Monitor
Wipe the exterior of the monitor with IPA (70% isopropyl alcohol), bleach
(100 ppm), or a mild detergent and water.
Cleaning the GlideScope
Cobalt AVL Cradle
Wipe the cradle with a standard hospital-grade surface cleaning product.
Содержание Cobalt AVL
Страница 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Страница 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Страница 26: ......
Страница 27: ...GlideScope Cobalt AVL System User s Manual...
Страница 28: ......
Страница 29: ...GlideScope Cobalt AVL System User s Manual...
Страница 30: ......
Страница 77: ......