Caution. Risk of permanent equipment damage.
Do not expose GlideScope Cobalt AVL Video Batons to
temperatures above 60 °C (140 °F). Do not disinfect the
GlideScope Cobalt AVL Video Batons using equipment such
as autoclaves, ultrasonic cleaners, or pasteurizers. Use of
such methods will cause permanent device damage and void
the warranty. Refer to Cleaning Section for a list of approved
cleaning procedures and products.
The technician should check for the following items:
• External damage
• Damage to the power supply
• Connectors and cable insulation integrity
To ensure patient safety, users should perform a routine inspection of
the GlideScope Cobalt AVL Video Baton before every use to ensure that
all endoscopic components are free of unintended rough surfaces, sharp
edges, protrusions or cracks.
If inspection reveals any faults in the components, contact Verathon
Medical Customer Care. All repairs must be performed by an authorized
Verathon Medical Service Center.
Monitor the color of temperature
indicator to avoid overheating the
Cobalt AVL Video Baton
Temperature indicator
The temperature indicator turns black if the GlideScope
Cobalt AVL Video Baton is heated above 60 °C.
A gray or white indicator does not indicate overheating.
Содержание Cobalt AVL
Страница 1: ...GlideScope Cobalt AVL System User s Manual and Quick Reference Guide 0900 2654 04 20...
Страница 2: ...GlideScope Cobalt AVL System Quick Reference Guide...
Страница 26: ......
Страница 27: ...GlideScope Cobalt AVL System User s Manual...
Страница 28: ......
Страница 29: ...GlideScope Cobalt AVL System User s Manual...
Страница 30: ......
Страница 77: ......