FIELD REPAIR/SERVICE
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 89 of 110
TROUBLESHOOTING GUIDE
OBSERVATION
POSSIBLE
PROBLEM
ACTION TO BE TAKEN
H. The probe from the
patient is inserted
but when operating
in
AUTO CONTROL
MODE
, the
PATIENT display
does not light up and
alarm doesn't sound.
(cont.)
Cable from the probe jack
to microprocessor board is
disconnected, SETPOINT/
STATUS display flashes
CHECK PROBE and the
alarm sounds.
Reconnect cable (J8 position on
microprocessor board).
Probe jack on side of unit
is defective, SETPOINT/
STATUS display flashes
CHECK PROBE or BAD
PROBE and the alarm
sounds.
Replace the probe jack.
Assembly
I. The probe from the
patient is inserted
but when operating
in
MANUAL
CONTROL MODE
,
the PATIENT display
does not light up and
alarm doesn't sound.
The patient temperature is
not within the appropriate
temperature range.
Check the location and
temperature of the probe.
The probe is broken or
defective.
Set the BLANKETROL II to
MONITOR ONLY MODE
to
determine if the probe is broken
or defective.
J. The probe from the
patient is inserted
but when operating
in
MONITOR ONLY
MODE
, the
PATIENT display
doesn’t light up and
alarm doesn't sound.
Other than a 400 Series
probe was inserted in the
probe jack.
Replace with the correct 400
Series probe.
The patient temperature is
not within the appropriate
temperature range.
Check the location and
temperature of the probe.