MAINTENANCE
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 61 of 110
F. Clean the wire mesh and the plastic cap. Be careful not to lose the black O-ring in the
rim of the plastic cap.
G. Replace the wire mesh in the plastic cap and position the mesh, O-ring and plastic cap.
H. Screw the cap clockwise onto the fitting until it is secure.
I. Reposition the front storage drawer as described in Section (5-1.6).
J. Replenish the water reservoir as described in Section (4-2.5.).
4-4. MAINTENANCE OF THE CONDENSER AND GRILL
Quarterly the grill and the fins of the condenser should be cleaned to remove any accumulated
dust and lint. To clean the condenser and grill:
A. Remove the grill.
B. Using a brush or a vacuum, dust both sides of the front panel grill.
C. Next, brush or vacuum the fins of the condenser. Be careful not to bend any of the fins.
D. Replace the grill.
4-5. MAINTENANCE OF THE BLANKETROL II EXTERIOR
– CLEANING INSTRUCTIONS
The BLANKETROL II unit is constructed of powder coated steel with a plastic top that should
be cleaned quarterly with water and mild detergent solution. For cleaning and disinfecting,
always use conventional hospital-approved topical equipment cleaners and disinfectants that
do not contain alcohol. Avoid alcohol and other strong, undiluted disinfectants. These may
cause staining of the device’s outer skin. Thoroughly wipe down device with a damp cloth to
remove any residue from cleaning solutions.
WARNING
The repair, calibration, and servicing of the BLANKETROL II should be performed
by qualified Medical Equipment Service Technicians, Certified Biomedical
Electronics Technicians, or Certified Clinical Engineers familiar with good repair
practices for servicing medical devices, and in accordance with instructions
contained in this manual.
Improper repair can result in damage to the
BLANKETROL II system and patient injury.
Always
unplug the unit
before accessing internal components during service.
Failure to unplug the unit could result in electric shock.
No modification of this equipment is allowed without prior, written authorization from
CSZ.
Failure to do so may result in damage to the BLANKETROL II system
and patient injury.
Before
returning the BLANKETROL II to patient use
after
repairs, the FIRST TIME
SET-UP/SYSTEM TEST ROUTINE in Section 2 must
always
be performed.
Improper repair and inadequate maintenance can result in damage to the
BLANKETROL II system and patient injury.