OPERATION
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 45 of 110
3-6. CONCLUDING HYPER-HYPOTHERMIA TREATMENT
Discontinue therapy upon the order of the physician. Patient's temperature can drift up or down
0.5°C ± 1°C (1°F ± 2°F) after therapy has been discontinued. The drift may be greater if the
patient has been shivering and treatment is abruptly discontinued. The operator should
continue to monitor the patient's temperature. To do so, the operator may choose to operate the
unit in
MONITOR ONLY MODE
as described in Section (3-5.).
To change the mode of operation or to stop the unit, the operator must press the TEMP SET
button or turn the power switch to
the “O” position. The operator can go directly to
MONITOR
ONLY MODE
from any of the control modes by pressing the MONITOR ONLY button.
When the hyper-hypothermia therapy is concluded and the unit is turned off:
A. Permit the blanket(s) and hose to remain connected to the unit for about ten minutes.
This allows some of the water to drain back into the unit.
B. Remove the probe from the patient and probe jack. Maintenance of the REUSABLE
probe is described in Section (4-7.).
C. Disconnect the power cord from the power source, loosely coil it and attach it to the back
panel using the nylon strap.
D. Remove the blanket(s).
E. For reusable PLASTIPAD blankets, loosely coil the hose lengthwise in the center of the
blanket. Fold the blanket lengthwise into the center, 1/3 from the left side and 1/3 from
the right side. Maintenance of the blanket is described in Section (4-6.).
For reusable Gelli-Roll blankets/pads either lay them flat or roll the blanket/pad. Do no
fold the Gelli-Roll blankets/pads.
For Single-Patient blankets, follow the instructions packaged with the blanket.
F. Disconnect the connecting hose from the unit and store in the front storage drawer.
3-7. SETPOINT/ STATUS DISPLAY MESSAGES
The SETPOINT/ STATUS display located in the center of the BLANKETROL II unit membrane
control panel reports the operating status of the unit or indicates changes the operator must
make. The SETPOINT/ STATUS display can show different messages to guide the operator.
The following list defines each message and describes the changes, if any, the operator must
make. Note that in the list below, the symbol “x” is used to indicate a numerical digit that can
change based on measurements or operator actions and “Y” is used when C or F would be
displayed depending on the temperature scale in use.