Warning
Remove the BLANKETROL II from service if the outer casing or membrane control panel is cracked
or internal components are exposed.
Contact with internal components could result in electric
shock or thermal injury to the patient or operator and exposure to sharp edges.
Keep grill and condenser clean and free of debris and obstruction.
Blockage of the grill and
condenser could result in the unit overheating which could render the unit unable to provide
adequate therapy and excessive surface temperatures could cause injury of patient or
operator. Keep unit, specifically the grill away from curtains or other obstructions.
Working with electronic boards, plugs, and cables requires delicate handling. Proper
electrostatic
discharge (ESD)
procedures should be followed during replacement of any electronic board.
Failure to do so may result in damage to the board.
Use of accessories other than those specified in the Operation and Technical Manual
may result in
increased electromagnetic emissions or decreased immunity to electromagnetic emissions
of the BLANKETROL II unit. This could affect the BLANKETROL II’s compatibility with other
electrical equipment. Electromagnetic compatibility refers to electronic devices unintentionally
affecting the operation of each other by emitting electromagnetic energy.