OPERATION
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 31 of 110
6. The heater or compressor may be activated.
7. The water flow indicator on the right side panel begins to move. The water moves
from the unit through the blanket and returns to the unit.
8. Unit operates in
AUTO CONTROL MODE
.
W. Press TEMP SET button
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The SETPOINT/ STATUS display returns to set temperature mode.
X. Press the MANUAL CONTROL button.
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The WATER display shows the temperature of the water in the reservoir.
4. The PATIENT display shows the temperature reading of the probe, as it is held by
the operator, if the probe reading is in the range of 10°C - 50°C (50°F - 122°F).
5. The SETPOINT/ STATUS display shows either:
XXXXXX WATER
MANUAL SETPT 37.0 C
Or SETPOINT/ STATUS display shows:
WATER @SETPT
MANUAL SETPT 37.0 C
NOTE
:
“XXXXXX” is HEATING, or COOLING.
The above depends upon the relationship of the blanket water temperature to
the set point temperature.
NOTE (ENGLISH MEMBRANE ONLY):
SET POINT
may be displayed in
Fahrenheit if the operator is in Fahrenheit mode.
6. The pump is activated.
7. The heater or compressor may be activated.
8. The water flow indicator on the right side panel begins to move.
This step is included to show that the patient display lights up when the probe is inserted
and the unit is in the
MANUAL CONTROL MODE
as well as when the unit operates in
AUTO CONTROL MODE.
Y.
Press TEMP SET button
1. The microprocessor board beeps.
2. The LED in the corner of the button lights up.
3. The SETPOINT/ STATUS display returns to set temperature mode.
Z. To complete this test routine, turn the power switch off.
1. The membrane control panel goes blank.
2. The light of the power switch goes out.
AA.
Remove the 400 Series probe from the probe jack.
AB.
Disconnect the power cord from the power source, loosely coil it and attach it to the
back panel using the nylon strap.
AC.
For reusable - PLASTIPAD - blankets, disconnect the connecting hose from the unit.
Loosely coil the hose lengthwise in the center of the blanket. Fold the blanket