OPERATION
BLANKETROL II, Model 222S
OPERATION AND TECHNICAL MANUAL
Page 39 of 110
Q. If a top hyper-hypothermia blanket is to be used, follow instructions in
step (J).
R. Connect the top blanket to the BLANKETROL II unit following the procedure described
in Section (2-2.1.).
S. If no top hyper-hypothermia blanket is used, cover the patient with a top sheet and/or
blanket. Patient preparation and bedside care are further described in Section (2-4.).
T. Choose which operating mode to use:
1. Operating in
AUTO CONTROL MODE
is described in Section (3-2.).
2. Operating in
MANUAL CONTROL MODE
is described in Sections (3-3.) and (3-4.)
3. Operating in
MONITOR ONLY MODE
is described in Section (3-5.).
NOTE: The maximum contact surface temperature is 41°C (105.8°F).
NOTE
:
If at any time, power is removed from the unit without actuating the power switch (I/O
switch) a power fail alarm will activate and the Power fail symbol LED located on the
membrane control panel will flash. To stop the alarm, restore power or push the SILENCE
ALARM button.
In the case that the unit is off for 5 seconds or less, the unit will resume the previous
operation mode. Otherwise if power is off longer the unit will go into power fail alarm.
3-2. OPERATING THE BLANKETROL II UNIT IN
AUTO CONTROL MODE
The BLANKETROL II unit can be set so that
it monitors the patient’s temperature and delivers
the maximum heating or cooling therapy in order to bring the patient’s temperature to a set point
determined by the operator. To do so, set the desired patient temperature in Celsius or
Fahrenheit, insert or attach the 400 Series probe in or on the patient and depress the AUTO
CONTROL button. The BLANKETROL II unit activates to heat or cool the water, to circulate the
water, and to control and monitor the change in the patient's temperature.
WARNING
A physician's order is required for setting blanket temperature and use of equipment.
At least every 20 minutes, or as directed by physician, check patient's temperature
and skin integrity of areas in contact with blanket; also, check the BLANKETROL
II’s
water temperature. Pediatric patients, temperature-sensitive patients with vascular
disease, surgical patients, diabetics and patients with Raynaud
’s Disease are at
greater risk for developing tissue injuries, and this should be considered when
selecting the temperature, duration of therapy and frequency of skin checks.
Notify
the physician promptly of any change in patient status in order to avoid serious
injury or death.
CAUTION
If device is set to operate in any Automatic mode, the device will shut down and
activate the CHECK PROBE alarm if the patient temperature falls below 30.0°C
(86°F). Patient temperature must rise above 30.0°C (86°F) before restarting the
device in Automatic mode. The device can be operated in Manual Control mode in
order to warm the patient above 30.0°C (86°F). Potential for interruption of therapy.