![GE Vivid 3 Pro Скачать руководство пользователя страница 95](http://html1.mh-extra.com/html/ge/vivid-3-pro/vivid-3-pro_service-manual_4262532095.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
3-32
Section 3-5 - Completing the Hardware Installation
3-5-5-2
Connecting the Ultrasound Unit to the Electrical Outlet
Note:
To help assure grounding reliability, connect to a hospital-grade or “hospital only” grounded
power outlet. If using the ultrasound unit with an external UPS system, follow all the grounding
and applicable safety standards as documented both in this manual and the external UPS
manufacturer’s manual. The external UPS system is to be considered the AC outlet.
1) Verify that the AC wall outlet is of the appropriate type.
2) Turn off the AC circuit breaker at the rear of the unit.
3) Plug the power cord connector into the AC input socket, and secure it in place using the attached
clip - see
Figure 3-19
4) Plug the other end of the power cord to the AC wall outlet. Allow sufficient slack so that the plug will
not be pulled out if the unit is slightly moved. The remaining length of the cord should be looped
and hung on the hook provided.
DANGER: To avoid the risk of fire, power to the system must be supplied from a separate,
properly rated outlet. It is recommended to use a dedicated power outlet. The power plug should
not, under any circumstances, be altered to a configuration rated less than that specified for the
current. DO NOT use an extension cord or adaptor plug. Refer to the
Electrical Requirements
section, on page 2-3
for more details.
Figure 3-19 Circuit Breaker and Power Cable on Back of Scanner
Mains Power Cable
Cable Clip
Circuit Breaker
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com