Chapter 3 - Installation
3-53
3) Verify that the following options are installed, as required:
•
DICOM Media Creator
is required to support DICOM storage to MOD or CD-R media.
•
DICOM Network Connectivity
is required to support DICOM storage over the network, for
example, to Proslov.
•
EchoNET Connectivity
is required to support an EchoPAC direct connection.
•
Excel Export
is required to support the ability to export measurements and reports to Microsoft
Excel files.
4) Enable the
DICOM printer
option, as required (under the
Services
sub-tab) - see
Figure 3-34
.
Figure 3-34 Setting up DICOM Print
5) Enable any other remote server, as required (for example, Services, My Computer, select
MPEGvueCDR
as shown in
Figure 3-35
).