Section 7-2 - Diagnostics
7-2-3-1
Diagnostic Menu Options
The
Diagnostic
menu includes the following options:
•
Software Version:
Displays the Vivid™ 3 ultrasound unit software version number and last
upgrade date.
•
:
Displays test results, as described in the
Common Service Interface
section, on
page 7-84.
•
:
Calibrates the Front End crate, and enables you to perform the display and video
grabbing test, as described in
Calibration
on page 6 - 14.
•
:
Tests each board in the Front End crate including the TR power supply, the LVPS
supply and temperatures, as described in the
Performing Front End (FE) Diagnostics
section, on
page 7-5.
•
:
Tests various back end hardware components, as described in the
Performing
Back End Diagnostics on the System
section, on page 7-68.
•
:
Monitors all low voltages, Tx voltages, temperatures and AC power input.
•
:
Performs a full sequence of Front End tests that do not require user intervention.
Auto Sequence is particularly recommended for detecting problems that are intermittent in nature,
since a test may be paused at any stage and any detected failures reviewed. This is described in
the
Auto Sequence
section, on page 7-6.
•
Description Area:
Displays a description of the function performed when each of the buttons is
selected. Trackball the cursor over a button to display a description in the
Description
area.
•
Report Button:
Displays a current diagnostics report.
•
Exit Button:
Closes the
Diagnostic
menu.