![GE Vivid 3 Pro Скачать руководство пользователя страница 288](http://html1.mh-extra.com/html/ge/vivid-3-pro/vivid-3-pro_service-manual_4262532288.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
7-44
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-7
Port Test
The Port test enables you to check the general logic control and ports logic.
Note:
You must reboot the system after performing the Port test.
1) Disconnect all probes.
2) Access the
F.B.
tab of the
FE Diagnostics
dialog box, as described in the
Accessing the Front Board
Assembly Options
section, on page 7-25.
3) Trackball to the
Port Test
button and press
Select
. The port test is performed and the results are
displayed in the
Comments
area.
4) Check the test results and proceed as follows:
•
If no errors were identified, proceed with the next test.
•
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
•
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration
section, on page 6-14, and then repeat the test.
•
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
Chapter 5 - Components
and Function (Theory)
. If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
5) Reboot the system.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com