![GE Vivid 3 Pro Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/ge/vivid-3-pro/vivid-3-pro_service-manual_4262532073.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
3-10
Section 3-3 - Receiving and Unpacking the Equipment
CAUTION:
When using sharp tools to open packing materials, take care to avoid cutting or damaging
the contents.
Note:
Unless otherwise specified, referenced items are shown in
Figure 3-6
on page 3-9.
1) Cut and remove the three securing steel strips (item 13).
2) Cut the adhesive tape along the center of the top of the cardboard shipping carton (item 14).
3) Remove all the fastening staples (item 15) and remove the two screws (item 16) so that the
cardboard carton (item 12) is free of the wooden platform (item 4), and the narrow flap is free.
4) Open the narrow side flap of the carton, remove the ramp (item 5) and attach it to the narrow side
of the wooden platform (item 4 in
Figure 3-7
below).
5) Remove the small boxes containing the accessories, probes and peripherals from the shipping carton.
6) Clear away the carton cover (item 12) from the wooden platform on which the ultrasound unit is
standing.
7) Cut and remove the antistatic sheet (item 6 in
Figure 3-7
, above) that is wrapped around the
ultrasound unit, taking care not to damage the antistatic cover (item 11 in
Figure 3-8
below)
covering the unit. Remove the antistatic cover.
Figure 3-7 Platform View of Shipping Carton
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com