Chapter 7 - Diagnostics/Troubleshooting
7-31
The
Report
dialog box contains the following information:
•
Test Result Message:
Displays the test result.
•
FB Elements:
Displays a graphic representation of the front board elements, as follows:
•
Each element is represented by a button (highlighted in green) that displays the element,
and the channel number that is associated with that element.
•
Each button is highlighted yellow when selected.
•
Each button is highlighted red if errors were detected for one or more elements.
•
Checkboxes:
Enable you to manipulate bad channels:
•
Show Only Bad Channels:
Select this checkbox to display only bad channels in the
Display
area.
•
Mark bad channel:
Select this checkbox to manually mark the currently selected channel
as bad.
6) Check the test result message and proceed as follows:
•
If one or more channels were identified as bad, trackball to the
Save Report
button and press
Select
, then contact a GE Medical Systems Service Representative.
•
If no channels were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to each of the channels in the
Display
area and press select.
•
Trackball to the
Mark bad channel
checkbox and press
Select
.
•
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
7) Trackball to the
Start Close test
button and press
Select
. The closed FLA jumper test is performed.
and the test results are displayed in the
Report
dialog box, as shown in
Figure 7-22
on page 7-40.