![GE Vivid 3 Pro Скачать руководство пользователя страница 192](http://html1.mh-extra.com/html/ge/vivid-3-pro/vivid-3-pro_service-manual_4262532192.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
5-18
Section 5-1 - Overview
5-4-3
MUX Board
5-4-3-1
General Description
The MUX Board has 64 identical receive channels, and performs the following functions:
•
Receiving the channels and connecting them to the BF Boards.
•
Expanding the low/high channels in CW mode.
•
TEE motion control.
•
Monitoring temperature and voltage measurements via the I²C.
•
FE Bus switch. To reduce the noise in the FE cards, the MUX switches off the FE Bus to the FB
during the data acquisition in CW mode.
•
Min/max AC input monitoring, every ten seconds.
Figure 5-12 MUX Board Diagram
Input Channels
Multiplex
TEE
Control
Monitor
Temp. Voltages
Fan Operation
FE Bus Switch
FE
Bus
FB
I C
Interface
2
FB
FB
BF
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com