4400_0060
enFlow Service Manual Rev. F 04/12
Page 13 of 36
Instructions for replacing the Controller/AC Power Pack clock battery.
1.
Turn Controller/AC Power Pack over.
2.
To remove the bottom, unscrew Pole Clamp screw; unscrew 4 screws shown below. Next, lift the cover.
3.
Power Assembly
a.
On the power supply assembly, unscrew 2 screws shown below (see
Diagram 5)
.
b.
Gently lift up assembly by holding pole clamp, and prop up with a small block or box (see
Diagram 6)
.
Diagram 1
Diagram 2
Diagram 4
Diagram 5
Diagram 6
Diagram 3