4400_0060
enFlow Service Manual Rev. F 04/12
Page 26 of 36
Fluid
Temp
Heater
Temperature LED on
Warmer
Display on Controller
Audible
Alarm
30 °C
Active
Blue Flashing
30°C Blue Flashing
No
31 °C
Active
Blue Flashing
31°C Blue Flashing
No
32 °C
Active
Blue Flashing
32°C Blue Flashing
No
33 °C
Active
Blue
33°C Blue
No
34 °C
Active
Blue
34°C Blue
No
35 °C
Active
Green
35°C Green
No
36 °C
Active
Green
36°C Green
No
37 °C
Active
Green
37°C Green
No
38 °C
Active
Green
38°C Green
No
39°C
Active
Green
39°C Green
No
40 °C
Active
Green
40°C Green
No
41 °C
Off
Green
41°C Green
No
42 °C
Off
Green
42°C Green
No
43 °C
Off
Yellow
43°C Yellow
No
44 °C
Off
Yellow
44°C Yellow
No
45 °C
Off
Red Flashing High Priority
Alarm after 20 seconds
45°C after 20 seconds
Red Flashing "Over
Temp" message
After 20
seconds
46 °C
Off
Red Flashing High Priority
Alarm after 16 seconds
46°C after 16 seconds
Red Flashing "Over
Temp" message
After 16
seconds
47 °C
Off
Red Flashing High Priority
Alarm after 12 seconds
47°C after 12 seconds
Red Flashing "Over
Temp" message
After 12
seconds
48 °C
Off
Red Flashing High Priority
Alarm after 8 seconds
48°C after 8 seconds
Red Flashing "Over
Temp" message
After 8
seconds
49 °C
Off
Red Flashing High Priority
Alarm after 4 seconds
49°C after 4 seconds
Red Flashing "Over
Temp" message
After 4
seconds
50 °C
Off
Red Flashing High Priority
Alarm (immediately)
Red Flashing "Over
Temp" message
(immediately)
Immediately
Appendix D:
Warming System Response by Temperature