4400_0060
enFlow Service Manual Rev. F 04/12
Page 12 of 36
Caution
Normal wear and tear during use of the Warmer may cause the device to be susceptible to fluid
ingress. Carefully inspect the heating surface of the Warmer for tears or foreign matter before each use
and take out of service if necessary.
For information on safety checks, please see Appendix F: Preventive Maintenance Procedure.
Controller/AC Power Pack (Model 120 series) - Setup Instructions
1.
Set the clock
a.
To modify the initial default mode of the Controller, press any of the front keys of the display
window prior to pushing the power switch that is located in the back of the unit. The key must
be held down continuously until the clock is displayed.
b.
Once the clock appears, toggle the center button to move through the fields. To change a
field, press the right or left arrow keys.
2.
Set the Controller/AC Power Pack display default mode to symbols
a.
Continue to toggle through to the set zone field. Press either the right or left arrow key to
access the padlock symbol to enter the password screen.
b.
An underscore will display under the first digit field. Press the arrow key to set the appropriate
digit.
c.
Next, press the center key again to toggle to the next place.
d.
Repeat steps three and four for the second and third digits. If an incorrect password is
entered, the system reverts to the set zone screen.
e.
After the third digit is set, press the center key again. First a green padlock and then the zone
USA will appear. Press the right arrow to change to INTL.
f.
Press the center key again to set. The system begins to operate.
g.
After the initial setup, whenever the Controller/AC Power Pack is powered on the display
screen defaults to the last mode entered.
h.
The password is 781.
Controller/AC Power Pack (Model 120 series) and AC Power Pack (Model 120
series)
The enFlow Controller/AC Power Pack (Model 120 series) and AC Power Pack (Model 120 series) contains
some parts that may be replaced. Check the fuses located in the power entry module if the Controller/AC
Power Pack or AC Power Pack fails to function. The AC line power cord must be removed to do this
.
Replaceable parts Controller/AC Power Pack and AC Power Pack:
Clock battery (See below for instructions on how to
replace.)
Fuse - Bussmann #GDB 5A or equivalent (See below
for instructions on how to replace.)
Cover (Bottom Only)
Screws
Pole clamp screw
Power cord
Warmer Mount
To order replacement parts, please reference the Recommended Supply List for specific product numbers.
For information on safety checks, please see Appendix F: Preventive Maintenance Procedure.