Although the Warmer has been tested to insure it will survive a drop of 1 m (3.28 ft.), care should be
taken that the device is not dropped to reduce the potential of damage.
Do not clean with:
o
ketones (MEK, acetone, etc.) or
o
abrasive cleaners.
Do not sterilize the Warmer with:
o
steam sterilization (autoclave) or
o
dry heat.
Do not disinfect or sterilize the Controller.
Do not spray or pour cleaning solutions directly on the Controller.
Do not allow cleaning solutions to accumulate on the Controller.
When using the Controller/AC Power Pack mounted to an IV pole, it must be tightly secured on the
pole no higher than 122 cm (48 in.) from the ground. The pole should have a base diameter of no less
than 61 cm (24 in.). A Controller/AC Power Pack mounted too high on the IV pole may cause
instability. IV pole accessories or the attachment of fluid bags may also cause instability.
Normal wear and tear during use of the Warmer may cause the device to be susceptible to fluid
ingress. Carefully inspect the heating surface of the Warmer for tears or foreign matter before
each use and take out of service if necessary.
Always secure the infusion set with the clip provided on the Warmer power cable to prevent kinking
in the line.
Do not use a stiff bristle brush or sharp probe to remove foreign material.
Do not use compressed air to dry.
Avoid puncturing the heating surface. If damaged, remove the Warmer from service and replace
immediately.
This equipment is not intended for use in an oxygen rich environment.
No modification of this equipment is allowed.
Do not position the device in a way which makes it difficult to disconnect the device.