Guidance and Manufacturer’s Declaration - Emissions
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment specified below. The
customer or user of the enFlow 100 with enFlow 120 should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
Radiated
CISPR 11
Group 1, Class B The enFlow 100 with enFlow 120 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Harmonics
IEC 61000-3-2
Class A
The enFlow 100 with enFlow 120 is suitable for use in all establishments,
other than domestic, and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic
purposes.
Flicker IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment specified below. The
customer or user of the enFlow 100 with enFlow 120 should ensure that it is used in such an environment.
Immunity Test
EN/ IEC 60601
Test Level
Compliance Level
Electromagnetic Environment – Guidance
ESD
EN/IEC 61000-4-2
Level 3 (±6kV)
Contact
Level 3 (±8kV)
Air
Level 3 (±6kV)
Contact
Level 3 (±8kV)
Air
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
r/h should be at least 30%
EFT
EN/IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
±1kV I/Os
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
EN/IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage
Dips/Dropout
EN/IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
100% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
See Note 1
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the enFlow
100 with enFlow 120 requires
continued operation during power mains
interruptions, it is recommended that
the enFlow 100 with enFlow 120 be
powered from an uninterruptible
power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be that of a typical commercial
or hospital environment.
Note 1) During the 5 Second event, the enFlow 100 and 120 power off, but return to normal operation as soon as
power is restored.
Guidance and Manufacturer’s Declaration – Emissions
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment
specified below. The customer or user of the enFlow 100 with enFlow 120 should ensure that
it is used in such an environment.
Immunity Test EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
EN/IEC 61000-
4-6
Radiated RF
EN/IEC 61000-
4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile communications
equipment should be separated from
the enFlow 100 with enFlow 120 by no
less than the distances calculated/listed
below:
D=(3.5/3V/m)(Sqrt P)
D=(3.5/3V/m)(Sqrt P)
80 to 800 MHz