Chapter 4. Inserting T17 Transducers into the Pipe
72
Model T17 Installation Guide
4.3 Connecting an XAMP (cont.)
8.
Place the
XAMP
body into the junction box (see
below), resting the puck gently on the cables below
it. Ensure that the cables of the
XAMP
rest naturally according to the slant at which they exit the epoxy, to
reduce stress and strain on the joint. The
XAMP
should remain still, and the cap of the junction box should rotate
freely around the
XAMP
.
Figure 86: Junction Box Cap
9.
Place the cover of the junction box over top of the
XAMP
and tighten the cover until the unit is closed and secure.
Engage the set screw to discourage tampering or removal (see
below).
Figure 87: Set Screw on Junction Box
Note:
To disconnect or uninstall the
XAMP
from the assembly, just reverse the procedures above.
Содержание T17
Страница 1: ...GE Oil Gas Flow 916 128 Rev D September 2020 Model T17 Ultrasonic Flow Transducer Installation Guide ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 62: ...Chapter 2 Installing the Pipe Nozzles 54 Model T17 Installation Guide no content intended for this page ...
Страница 64: ...Chapter 3 Installing an Isolation Valve 56 Model T17 Installation Guide no content intended for this page ...
Страница 88: ...Chapter 6 Specifications 80 Model T17 Installation Guide no content intended for this page ...
Страница 90: ...Warranty 82 Model T17 Installation Guide no content intended for this page ...
Страница 92: ......
Страница 93: ......