Model T17 Installation Guide
35
Chapter 2. Installing the Pipe Nozzles
2.3.7 Hot Tapping the Pipe - MR1 Mid-Radius
For 3 in. Flanges:
1.
Install 3 in. ANSI flanged isolation valves on both nozzles (full bore, 8 in. face to face for 150# RF or
11.125 in. for 300# RF) with a gasket and 5/8 in. (for 150# rated flanges) or 3/4 in. (for 300# rated flanges)
diameter studs and nuts. Orient the valve handles to minimize interference.
2.
Hot tap holes in the pipe using a hot tap machine equipped with a 3/4 in. (19.05 mm) drill bit. Then use a
coupon retaining hole saw to cut a hole with a diameter of 2.50 in. (63.5 mm) minimum.
2.3.8 Cold Tapping the Pipe - MR1 Mid-Radius
The procedure for cold tapping a pipe is the same as the hot tapping procedure. However, an isolation valve is not
necessary during the process. The hot tapping medium can be applied directly to the nozzle. Isolation valves will be
added after the tapping is complete.
WARNING!
Hot tapping should be performed only by qualified personnel. Follow all applicable code
and safety practices during these procedures.
WARNING!
Cold tapping should be performed only by qualified personnel. Follow all applicable code
and safety practices during these procedures.
Содержание T17
Страница 1: ...GE Oil Gas Flow 916 128 Rev D September 2020 Model T17 Ultrasonic Flow Transducer Installation Guide ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 62: ...Chapter 2 Installing the Pipe Nozzles 54 Model T17 Installation Guide no content intended for this page ...
Страница 64: ...Chapter 3 Installing an Isolation Valve 56 Model T17 Installation Guide no content intended for this page ...
Страница 88: ...Chapter 6 Specifications 80 Model T17 Installation Guide no content intended for this page ...
Страница 90: ...Warranty 82 Model T17 Installation Guide no content intended for this page ...
Страница 92: ......
Страница 93: ......