page 23
© 2003 General Electric
THE AED HAS THREE MODES:
Operating Mode:
Defined as having the battery installed and the lid open. This is the mode the AED would be
in during an actual rescue situation.
Standby Mode:
When the battery is installed, but the lid is closed. In this mode the AED is not being used in a
rescue. The device will conduct its routine self-tests to ensure proper operation.
Storage Mode:
When the battery is removed, such as during shipping or transport. With the battery removed,
the AED is unable to perform self-tests or rescues.
ENVIRONMENTAL OPERATING AND STANDBY CONDITIONS
See Section 7 – Technical Data, Parameters, Environmental Operation and Standby Conditions on page 48.
CAUTION:
Temperature/Humidity/Pressure Extremes
Exposing the AED to extreme environmental conditions outside of its operating parameters may
compromise the ability of the AED to function properly. The RescueReady
®
daily self test verifies the
impact of extreme environmental conditions on the AED by checking temperature, humidity and
pressure; if the daily self test determines environmental conditions outside of the AED’s operating
parameters for 5 consecutive days, a "SERVICE REQUIRED" alert will be issued to prompt the user
to move the AED to environmental conditions within the acceptable operating parameters at once.
See Section 7 – Technical Data, Parameters, Operation and Standby Conditions on page 48.
SHIPPING AND TRANSPORT CONDITIONS
(For up to 1 week)
See Section 7 – Technical Data, Safety and Performance Standards/Shipping and Transportation Conditions on
page 52.
Section 3 - Getting Started
201-9198-001 Rev A