page 53
© 2003 General Electric
ENERGY LEVELS AND PATIENT IMPEDANCE
The Biphasic Truncated Exponential (BTE) waveform utilizes variable energy. The actual energy delivered will
vary with the patient’s impedance and the device will deliver a shock when impedance is between 20-180
Ohms. Energy will be delivered at three different levels referred to as ultra-low variable energy, low variable
energy, and high variable energy as shown in the above waveform tables. The accuracy of the energy for the
energy in a 50 Ohm resistor is ±15%.
Section 7 - Technical Data
Patient’s
Impedance
(Ohms)
25
50
75
100
125
150
Phase 1
Delivered
Start Volts
1609V
1640V
1651V
1656V
1660V
1662V
Phase 1
Delivered
End Volts
858V
1050V
1124V
1163V
1186V
1201V
Phase 1
Duration
(ms)
3.3 ms
4.5 ms
5.8 ms
7.0 ms
8.3 ms
9.5 ms
Phase 2
Delivered
Start Volts
858V
1050V
1124V
1163V
1186V
1201V
Phase 2
Delivered
End Volts
247V
553V
728V
835V
907V
959V
Phase 2
Duration
(ms)
3.2 ms
3.2 ms
3.2 ms
3.2 ms
3.2 ms
3.2 ms
Total Energy
Delivered
(J)
226J
200J
182J
169J
160J
153J
Table A2 - Low Variable Energy (200VE) Waveform into Different Resistive Loads (Typical Values)
Patient’s
Impedance
(Ohms)
25
50
75
100
125
150
Phase 1
Delivered
Start Volts
1869V
1906V
1918V
1925V
1928V
1931V
Phase 1
Delivered
End Volts
997V
1220V
1306V
1351V
1378V
1396V
Phase 1
Duration
(ms)
3.3ms
4.5ms
5.8ms
7.0ms
8.3ms
9.5ms
Phase 2
Delivered
Start Volts
997V
1220V
1306V
1351V
1378V
1396V
Phase 2
Delivered
End Volts
287V
643V
846V
971V
1054V
1114V
Phase 2
Duration
(ms)
3.2 ms
3.2 ms
3.2 ms
3.2 ms
3.2 ms
3.2 ms
Total Energy
Delivered
(J)
305J
270J
246J
229J
216J
207J
Table A3 - High Variable Energy (270VE) Waveform into Different Resistive Loads (Typical Values)
201-9198-001 Rev A