These protocols are selected by using our MDLink software program. The five biphasic energy protocols
available are as follows:
Energy
Shock Energy
Protocols
Sequence
1
Level
Energy Range (J)
Factory Default
1.
200VE
140J-250J
2.
300VE 190J-360J
3.
300VE 190J-360J
Protocol #2
1.
200VE
140J-250J
2.
200VE 140J-250J
3.
300VE 190J-360J
Protocol #3
1.
150VE
105J-195J
2.
200VE 140J-250J
3.
200VE 140J-250J
Protocol #4
1.
150VE
105J-195J
2.
150VE 105J-195J
3.
200VE 140J-250J
Protocol #5
1.
200VE
140J-250J
2.
200VE 140J-250J
3.
200VE 140J-250J
1
The ultra-low current, low current and high current shocks are variable energy. The actual energy is determined by the patient’s impedance.