page 31
© 2003 General Electric
Section 3 - Getting Started
VOICE PROMPT
“Stand Clear! Push
Flashing Button to
Deliver Shock.”
“Check for Breathing. If Not
Breathing, Give Patient Two
Breaths.”
“Check for Signs of
Circulation. If No
Circulation, Start CPR.”
“Check Electrodes”
“Battery Low”
“Analysis Interrupted. Stop
Patient Motion.”
“Open Lid to Continue Rescue”
“Rhythm Changed.
Shock Cancelled.”
“Data in memory. Push
Flashing Button to Erase
Data and Perform Rescue.”
“Remove Cable to
Continue Rescue.”
TEXT DISPLAY
STAND CLEAR
PUSH BUTTON TO SHOCK
IF NOT BREATHING GIVE
TWO BREATHS
IF NO CIRCULATION, START
CPR
CHECK ELECTRODES
BATTERY LOW
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
OPEN LID TO CONTINUE
RESCUE
RHYTHM CHANGED.
SHOCK CANCELLED
DATA IN MEMORY
PUSH BUTTON TO ERASE
REMOVE CABLE
SITUATION
After the AED is fully charged and ready to deliver the
defibrillation shock. The RED
SHOCK
indicator flashes
and the phrase repeats for 30 seconds or until the
SHOCK
button is pushed.
• After the AED delivers 3 consecutive defibrillation
shocks
• After the AED detects a non-shockable cardiac rhythm.
• When 2 1/2 minutes or more has elapsed since CPR
was last administered
• Note: This prompt may be disabled in MDLink,
• After previous prompt or if previous prompt is disabled
• After the AED delivers 3 consecutive defibrillation
shocks
• After the AED detects a non-shockable cardiac rhythm
during cardiac rhythm analysis
• When 2 1/2 minutes or more has elapsed since CPR
was last administered
Occurs when patient impedance is too low or too high.
Occurs once when the battery voltage becomes low,
although a rescue can continue for approximately 9
more shocks. When the battery is too low to do a
rescue, the phrase repeats continuously. You must
replace the battery before continuing with the rescue.
If completely depleted, all AED activity will terminate.
When the AED detects ECG noise artifact, stop moving
or touching the patient. Remove other electronic
devices within a 5-meter radius.
When the lid is inadvertently closed during a rescue,
this prompt will repeat for 15 seconds.
When the AED detects a change in rhythm – when the
device is prepared to shock.
When there is a previously stored rescue in internal
memory, the yellow
CONTINUE
button will flash and the
phrase will repeat.
The message can be cleared by pressing the
SHOCK/CONTINUE
button to erase the internally stored
rescue data or retrieving the rescue data with
RescueLink and erasing the stored data with
RescueLink. When a serial communication cable is
connected to the AED during a rescue, the phrase
repeats until the cable is disconnected.