page 46
© 2003 General Electric
201-9198-001 Rev A
8.
Q:
When I open the lid, why do I get the voice prompt “Data in Memory. Push Flashing Button to Erase
Data and Perform Rescue.” How to I get the message to stop occurring?
A:
This message occurs when there is a previously stored rescue in the internal memory of the AED.
You can clear the message by:
1. Pressing
the
SHOCK/CONTINUE
button to erase the internally stored rescue data OR
2.
Retrieving the rescue data with RescueLink and erasing the stored data with RescueLink.
9.
Q:
The AED did not sound an audible alert when I removed the electrodes and closed the lid. Why?
A: The lid-closed electrode self-test only activates the
STATUS INDICATOR
. The AED allows time for
replacement of the electrodes – as removing electrodes is a normal procedure after a rescue - or a
battery during the post rescue procedure, however, an audible maintenance indicator will be
triggered after the next Daily Self-Test.
10. Q:
What can I do to keep the AED warm when a rescue is in an isolated area and at subzero temperatures?
A:
When travel to a rescue involves exposing the AED to extremely cold temperatures for an extended
period of time, keep the electrodes and the battery warm.