page 49
© 2003 General Electric
ELECTRODES
•
Self-adhesive, disposable defibrillation electrodes
•
Minimum combined surface area: 228cm
2
•
Extended length of lead wire: 1.3m
LITHIUM BATTERY SPECIFICATIONS
•
Output voltage: 12VDC (max)
•
Batteries are non-rechargeable
•
Lithium contents: 9.2g (max)
•
Check local regulations for disposal information
Estimated
Typical
Model
Shelf Life
Warranty
Shocks
9142 Lithium
5 Years
3 Years
up to 290 shocks
The battery operating life depends on the type of battery (9142 for Responder AED), actual usage and
environmental factors.
BATTERIES AND CAPACITOR CHARGE TIMES
A new battery typically takes 10 seconds to charge the AED to maximum energy.
A battery with reduced capacity causes the red LED light to initially turn ON and typically takes 13 seconds to
charge a fully discharged AED to maximum energy.
The maximum time from “Power On” to “Ready to Shock” is 28 seconds for a new rescue.
The maximum time from “Analyze” to “Ready to Shock” is 22 seconds for a new rescue.
DELIVERY OF THREE DEFIBRILLATION SHOCKS
55 seconds (nominal)
Section 7 - Technical Data