page 51
© 2003 General Electric
EMISSIONS
IMMUNITY
ENVIRONMENTAL CONDITIONS
Section 7 - Technical Data
Field
E-M
Magnetic
Models
9300
9300
Standard or Compliance
IEC 55011/CISPR 11, Group 1, Class B
ANSI/AAMI DF39, <0.5mT on surface, except for within
5cm of the lid magnet and the speaker
Field
E-M
Magnetic
ESD
Models
9300
9300
9300
Standard or Compliance
IEC 61000-4-3, Level X, (20V/m)
IEC 60601-2-4, Section 36.202.3 (20-V/m)
AAMI DF39, Section 3.3.21.2.1
IEC 61000-4-8 (2001)
IEC 60601-2-4 (2002), Section 36.202.8
AAMI DF39, Section 3.3.21.2.3 80A/m, 47.5Hz – 1,320Hz
IEC 61000-4-2, Level 3
IEC 60601-2-4 (2002), Section 36.202.2
6KV contact discharge, 8KV air gap discharge
Condition
Free Fall Drop
Bump
Vibration (Random)
Vibration (Sine)
Enclosure Protection
Models
9300
9300
9300
9300
9300
Standard or Compliance
IEC 60068-2-32 (1975) AM 2 (1990), 1 meter
IEC 60068-2-29 (1987), 40g and 6000 bumps
IEC 60068-2-64 (1993): 10Hz –2KHz, 0.005 – 0.0012 g
2
/Hz
IEC 60068-2-6 (1995): 10Hz – 60Hz, 0.15 mm and 60Hz –
150 Hz, 2g
IEC 60529 (2001), IP24
201-9198-001 Rev A