4 - 22
Section 4-3 - General Procedure
4-3-10
Basic Measurements
NOTE:
The following instructions assume that you first scan the patient and then press Freeze
.
4-3-10-1
Distance and Tissue Depth Measurements
1.) Press
Measure
once, an active caliper displays.
2.) To position the active caliper at the start point (distance) or the most anterior point (tissue
depth), move the
Trackball
.
3.) To fix the start point, press
Set/ B Pause
. The system fixes the first caliper and displays a
second active caliper.
4.) To position the second active caliper at the end point (distance) or the most posterior point
(tissue depth), move the
Trackball
.
5.) To complete the measurement, press
Set
. The system displays the distance or tissue depth
value in the measurement results window.
Before you complete a measurement:
To toggle between active calipers, rotate
Cursor Select
button.
To erase the second caliper and the current data measured and start the measurement again, press
Clear
.
NOTE:
To rotate through and activate previously fixed calipers, rotate Cursor Select button.
NOTE:
After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press Clear.
4-3-10-2
Circumference/Area (Ellipse) Measurement
1.) Press
Measure
once; an active caliper displays.
2.) To position the active caliper, move the
Trackball
.
3.) To fix the start point, press
Set/ B Pause
. The system fixes the first caliper and displays a
second active caliper.
4.) To position the second caliper, move the
Trackball
.
5.) Rotate the
Ellipse
button; an ellipse with an initial circle shape appears.
NOTE:
Be careful not to press the Ellipse control as this activates the Zoom.
6.) To position the ellipse and to size the measured axes (move the calipers), move the
Trackball
.
7.) To increase/ decrease the size, Rotate the
Ellipse
button.
8.) To toggle between active calipers,
rotate Cursor Select button
.
9.) To complete the measurement, press
Set/ B Pause
. The system displays the circumference
and area in the measurement results window.
Before you complete a measurement:
-
To erase the ellipse and the current data measured, press
Clear
once. The original caliper is
displayed to restart the measurement.
-
To exit the measurement function without completing the measurement, press
Clear
a second
time.
4-3-10-3
Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
Содержание LOGIQ V3
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3 ...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO ...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks ...
Страница 145: ......
Страница 169: ......
Страница 170: ... 2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...