GE
LOGIQ V5/LOGIQ V3
D
IRECTION
5496012-100, R
EVISION
3
B
ASIC
S
ERVICE
M
ANUAL
4 - 28
Section 4-3 - General Procedure
4-3-13-4
Archiving Images
1.) Insert the archive media.
2.) To format the archive media, enter
Utility
->
Connectivity
->
Removable Media
.
3.) Format the CD. Verify the format if desired.
4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by
date.
NOTE:
Images will be moved from the hard drive by date. Therefore, the best way to label media is by
date. When images are moved to the archive media, they will be deleted from the system hard
drive. However, the patient database (backed up earlier) maintains pointers to the location of
the images on the archive media.
5.) Enter
Utility
->
System
->
Backup/Restore
,
6.) Select “Move File Older Than in Days”.
Figure 4-20 Format Media Screen
Figure 4-21 EZBackup/Move